Ground fault circuit interrupter test – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 96

Advertising
background image

Product Service 1-800-522-7658

96

Ground Fault Circuit Interrupter Test

(115/100 Volt models only)

Note: The GFCI tester must be able to simulate a fault of 3mA.

1.

Place the tester into any of the Logic+’s outlets. The indicator lights of the tester

should indicate normal function, and the GFCI’s indicator LED should be lit.

2.

Press the test button of the tester. The indicator lights should now indicate the

circuit is inactive, and the GFCI indicator LED should be off.

3.

Reset the GFCI by pressing the RESET button on the right outlet. The tester

should again indicate normal operation, and the GFCI indicator LED will

illuminate.


Figure 5-24





GFCI Reset Button

GFCI Test Button

GFCI Indicator LED

Acceptance

The GFCI trips during the simulated failure, and the indicator LED turns off. Pressing

the Reset button returns the GFCI to normal function.