Vacu-pass, Cord & cable portal installation – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 154

Advertising
background image

Product Service 1-800-522-7658

154

Vacu-Pass

TM

Cord & Cable Portal Installation

Note: You will need to access the right side of the cabinet in order

to install the Portal
Note:
There must be enough clearance to pass the cord or cable

between the exterior dress panel and any obstruction.
Note: Some Vacu-Pass components and the cord or cable passing

through it may become contaminated during use of the cabinet.

Ensure all potentially contaminated components are surface

decontaminated before handling or removal from the cabinet.
1.

Open the right side panel by removing the two Phillips screws

on the side panel.

2.

Remove the four locknuts that secure the cover panel to the

liner. Keep the nuts for future use.

3.

Using a putty knife or similar device, carefully pry the cover

away from the liner. Leave the sealing gasket in place.

4.

Install the body of the portal into the liner, and sequentially

tighten the nuts unto a uniform gasket compression is noted.

5.

Remove the grommet from the liner side wall. Remove the

solid sealing plug from the body of the portal by either

pressing it through from the outside, or by carefully inserting

a spatula or similar device between the sealing plug and the

body of the portal, and prying the plug out.

6.

Pass the cord or cable through the body of the portal, and

then through one of the plugs that has been cut for cord or

cable use, then through the grommet, as shown.


Note: select a plug with a hole

that is slightly smaller than the

cord or cable, to create a

proper seal. This will also help

minimize movement of the cord

or cable if it is accidentally

pulled during use.

7.

Position the cord or cable as

it will be used in the

cabinet, and then push the

plug back into the body of

the portal until it seats in

the portal. Reinstall the

grommet

.