Labconco PURICARE Procedure Station Models 31260 User Manual

Page 167

Advertising
background image

Product Service 1-800-522-7658

167

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower runs

slower than

normal; speed

control has little

effect on blower

speed. Blower may

take more than 5

seconds to start.

Motor signal wires

damaged or

disconnected

Measure motor output signal

voltage. Inspect signal wiring and

connections

Power Supply

Board

misconnected or

defective

Test output voltages on the Power

Supply board.

Motor jumper wire

is damaged/

disconnected

(115/100 Volt

models only)

Inspect grey wire #31 on 15 pin

connector for continuity.

Blower motor is

defective

Replace blower motor.

Keypad

disconnected or
defective

Run keypad diagnostics and check

connections.

Fluorescent light

not working

Sash is closed

Open sash – Fluorescent lights will

not work with the sash closed.

Lamp(s) are

defective

Replace defective lamp(s)

Lamp wiring is

disconnected

Inspect lamp wiring.

Defective lamp

ballasts

Replace lamp ballasts.

Keypad

disconnected or

defective

Run keypad diagnostics and check

connections.