Labconco PURICARE Procedure Station Models 31260 User Manual

Page 108

Advertising
background image

Product Service 1-800-522-7658

108

Figure 6-19





Exhaust HEPA Filter bolts




4.

If you choose to remove the exhaust filter pan, remove the sealing bolts, lift the

pan straight up, and then pull it forward and out, as shown in Figure 6-20.

5.

When reinstalling it, ensure all four edges of the pan are OUTSIDE the sealing

edges of the plenum, as shown in Figure 6-21.

Figure 6-20


Figure 6-21

Plenum sealing edges (inside the

exhaust filter pan on all 4 sides)