Electrical requirements – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 48
Product Service 1-800-522-7658
48
Electrical Requirements
The biosafety cabinet models have the following electrical requirements:
Model #
Requirements
All 3-& 4-foot Type A ending in -00
115 VAC, 60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -10
100 VAC, 50/60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -20,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All Type B2 ending in -00
115 VAC, 60 Hz, 12 Amps
All Type B2 ending in -10
100 VAC, 50/60 Hz, 12 Amps
All Type B2 ending in -20, 30, 40,
50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All 5-& 6-foot Type A ending in -00
115 VAC, 60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -10
100 VAC, 50/60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -20,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
Note: A dedicated outlet with an appropriate circuit breaker should
be located as close as possible to the right rear side of the cabinet,
at a height even with, or higher than, the top of the bench or stand.
Consult your local electrical codes for properly rated circuit
breakers. For safe operation the dedicated outlet must provide the
protective earthing ground connection to the cabinet.
Note: On 100 and 115 VAC models, both electrical outlets are protected by
a ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.
Do not use any detachable power cord that is not adequately rated
for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
Table 2-1