Section 5

Certification Procedures for the Logic+


This section will outline test procedures to validate the performance of the Logic+.

Note: These are the only Labconco-approved test procedures for validating

the performance of the Logic+ Biosafety Cabinet. Other test procedures may

yield different or inaccurate test results.
NOTE: In order to properly certify the Logic+, you must obtain a Logic+

Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by
contacting Labconco Customer Service Department at 800-821-5525
.
Note: The ECM motor is programmed by Labconco to deliver a constant

volume of air throughout the life of the HEPA filters. Any changes to the

blower speed control should be small (less than 5 units).

Accessing the Certifier’s Menu

Note: You will need to access this menu to do any of the following:

Adjust the blower speed

To reset the filter gauge

To calibrate the optional airflow sensor

Adjusting the blower speed

1.

With the Logic+ operating, press: “

Menu

2.

The Display will change to Menu mode. Press the

or

arrows until “

Service

is selected. Press: “OK/Mute”

3.

The Display will then change to the password mode. In sequence, press:

“Light”

“UV Light”

“Timer”

“Timer”

“OK/Mute”

4.

NOTE: The cabinet blower will automatically stop for approximately 60

seconds to zero the unit, and then automatically restart the blower. The
display will change to the speed setting mode. The number displayed is the

speed control setting, in this case a pulse width modulation (PWM) signal that

indicates the percentage of the motor’s duty cycle. The number should range

from 60 to 90 depending on the Logic+ model. To increase or decrease the

blower speed, press the

or

buttons, respectively. When the speed control

is set properly, press “

OK/Mute

” This will set the controller to its new value,

and change the display to the filter gauge mode.