Alarm Screens

Sash is too high

The sash is open too far for safe operation.

Airflow Alert

The airflow patterns in the cabinet have changed,

resulting in a sudden change in the motor speed.
This is most likely due to a blockage of the grille or

the exhaust filter outlet. It may also be caused by

removal of the work surface while the cabinet is in

operation.

Airflow Alarm

If equipped with the optional airflow sensor and

the inflow or downflow velocities are excessively

high or low, this alarm will be displayed.

Canopy Alarm

If equipped with the optional canopy connection,

and there is insufficient exhaust system airflow,

this alarm will be displayed. If this alarm is

displayed, the Logic+ blower will need to be

turned off, and then back on again to reset the

alarm.

System Error

The motor and display circuit board are not
communicating properly. DO NOT USE THE

CABINET UNTIL THE PROBLEM HAS BEEN

CORRECTED.