Labconco PURICARE Procedure Station Models 31260 User Manual

Page 142

Advertising
background image

Product Service 1-800-522-7658

142

Air Flow Sensor Kit No. 3405500

Note: A qualified certifier must install and calibrate this sensor to obtain

accurate operation.

Note: Exercise caution when working on the exhaust filter cover, to prevent

damaging the Exhaust HEPA filter.

Installation Procedure-Units without the Ventus Canopy Connection

1.

Disconnect power to the Biosafety Cabinet.

2.

Loosen the two Phillips screws on the front sides of the

exhaust cover, as shown

3.

Tilt the exhaust cover up to access the

underside of the cover.

4.

Remove the two screws that hold the airflow

sensor into its body, as shown. DO NOT pull

the sensor out of the body.

5.

Position the sensor so that the

screen on the sensor body is

oriented towards the front of

the cover on the underside of

the exhaust cover, as shown.

Insert and tighten the screws

from the top side of the cover.