Labconco PURICARE Procedure Station Models 31260 User Manual

Page 130

Advertising
background image

Product Service 1-800-522-7658

130

Figure 7-7

15 pin connector

4.

Disconnect the 15-pin connector, and install the pass-thru harness, as shown

in Figure 7-8. Plug the Logic+ back in. With the meter set on AC voltage,
verify the AC mains line voltage is present within ± 5 VAC when the AC

voltage is measured across pin 2 (black wire) and pin 3 (white wire) on the

3-pin connector, as shown in Figure 7-9.

Figure 7-8

Pass-thru Harness installed

Figure 7-9

Measuring Line voltage


5.

Set the meter to read DC voltage and connect the –DC voltage lead to pin 8

(brown wire) on the 4-pin connector. This brown wire is circuit common and

will be used for the following low voltage tests.