Type A HEPA Filter Replacement

CAUTION: The cabinet must be properly decontaminated before servicing the

supply HEPA filter. Following replacement of a HEPA filter, the cabinet must

be recertified.

CAUTION: Always remove the exhaust HEPA filter first. The exhaust filter pan

or plenum can be damaged if the plenum is raised with the Exhaust HEPA

filter in place.

CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate

lifting techniques to remove and handle it. Use two people to remove the

larger filters if possible.

To Remove the Exhaust HEPA Filter on 4-, 5- and 6-foot models:
1.

Unplug the cabinet.

2.

Remove the front dress panel as shown in Figure 6-1.

3.

Loosen all of the Phillips screws on the blower chamber cover. Remove all of the

screws except two screws on the top center of the panel. This will support the panel.

4.

Release the damper seal by pushing the outside edge of the seal towards the bolt

and inside the blower plenum cover, as shown in Figures 6-14 and 6-15.