Towel Catch Removal

Although not normally required, the towel catch can be removed for cleaning,

inspection, etc.
Note: The work surface of the cabinet and the towel catch must be thoroughly

decontaminated before removing either.
Although not normally required, the towel catch can be removed for cleaning,

inspection, etc.
Note: The work surface of the cabinet and the towel catch must be thoroughly

decontaminated before removing either.
1.

Remove the work surface as described above.

2.

Remove the towel catch by pivoting the bottom out toward you, as shown in Figure

6-32; it is spring loaded, and you will feel some resistance. Note the orientation of

the towel catch.

3.

Surface decontaminate the towel catch before removing it.

4.

Reinstall the towel catch by sliding it back into position, in the correct orientation.

Also ensure that the Sampling/Decontamination Tube (the black tube with an orange

cap) rests behind the towel catch, allowing the catch to contact the rear wall of the

cabinet.

Figure 6-32



Sampling/Decontamination

Tube

Towel Catch Retaining

Spring

Towel Catch

Note: Upper and lower

towel catch flanges point to

the front of the cabinet, and

the bottom of the towel

catch rests against the back wall.