Work Area Air Cleanliness Test

(optional)

If the customer wants to verify air cleanliness in the work area, use the following

procedure:

1.

Mark test points at the geometric center of the work surface, and midway

between the center test point and each side wall.

2.

Place a particle counter pickup a test point, at a height of 16 inches above the

work surface, pointing into the airflow, as shown in Figure 5-23. Take three

samples of 1.0 ft

3

. Repeat for each test point.

3.

Average the results.

Figure 5-23




Acceptance

The average count should be less than 100 particles 0.5um in size per ft

3

.