4.

Open the following number of Laskin nozzle(s):

For all 3-foot Logic+ models – 1 nozzle

For all 4-, 5- and 6-foot Logic+ models – 2 nozzles

5.

Ensure the pressure gauge on the generator is reading 20 +/- 1 PSIG.

6.

Scan the downstream side of the HEPA filter by passing the sampling nozzle of

the gun in slightly overlapping strokes over the entire surface of both of the

filters, with the sampling port not more than 1 inch from the surface of the filter

media. Scan the entire periphery of the filters and the gaskets between the filter

frame and the Logic+ frame. Scanning shall be done at a traverse rate of not

more than 2 inches per second.

Note: Operation of the photometer may become erratic when testing the front edge of

the supply filter frame, due to aspiration of room air into the front of the work area. This

problem can be minimized or eliminated by placing the edge of a sheet of rigid plastic or
metal at the edge of the filter gasket when scanning this area, as shown in Figure 5-5.

Figure 5-5

Acceptance

Aerosol penetration shall not exceed 0.01%.

Reassemble the unit by compressing the exhaust filter cover while installing its tabs

back into the exhaust filter bracket slots. Reinstall the diffuser, ensuring that the holes in

the back of the diffuser frame engage the posts on the rear wall of the work area, as

shown in Figure 5-6.

Figure 5-6

Diffuser pins engaging the holes in the rear of
the diffuser.