Labconco PURICARE Procedure Station Models 31260 User Manual

Page 147

Advertising
background image

Product Service 1-800-522-7658

147

4.

Carefully remove any remaining sealant from the service fixture hole.

5.

Using the PTFE tape supplied, wrap two to three turns of tape on the male pipe

thread of the valve, and the tube fitting.

6.

Put the threaded end of the fixture through the hole, orienting the valve as shown in

Figure 10-1. Using the silicone sealant, apply a bead of silicone on the outside

surface of the cabinet side, around the valve body. This will provide a pressure tight

seal around the valve body. Place the washer on the valve thread, and then install

the coupler. Holding the valve body in position, use a wrench or pliers to thread the

coupling on until the valve is captured on the side panel.

7.

Thread the tubing connector onto the coupling until it can no longer be turned by

hand. Holding the coupling secure with a wrench or pliers, turn the tube fitting

approximately one more revolution, using pliers or a wrench. Orient the tube fitting
such that it is easy to access it with the incoming service tube.

8.

Connect the ¼-inch O.D. service tube to the tube fitting by pressing the tube all the

way into the fitting, turning the tubing nut until hand tight, and then approximately

one more turn using pliers or a wrench.



Do not route the service tubing under the sash weight channel. Route the tubing

between the weight channel and the exterior side panel.

9.

Make sure the service fitting in the Purifier is in the off position before pressurizing

the service line. Once pressurized, check the valve body, the coupling, and the tube

fitting for leaks. Tighten any leaking joints.

10.

Reattach the side panel.