Labconco PURICARE Procedure Station Models 31260 User Manual

Page 143

Advertising
background image

Product Service 1-800-522-7658

143


6.

If you are installing the airflow sensor into a

canopy connection, install the bracket onto

the airflow sensor as shown.


7.

Gently remove the Canopy Cable Sealing grommet, located on

the top of the canopy. Remove the solid square plug. Locate

the replacement plug in the air sensor kit, and push the

sensor wiring harness through the slit in the plug.

NOTE: The grommet and plugs are keyed – they only fit

together properly when oriented as shown in the illustration.

Ensure both the replacement plug and the canopy are oriented

properly in the grommet before installation.

Route the sensor cable from the sensor, through the

grommet, and outside of the canopy. Reinstall the grommet.

Ensure the cable does not hinder the operation of the front

relief valve, and does not touch the surface of the HEPA filter.

8.

Then install the

sensor and canopy

bracket as shown.