Quickstart


How do I run a diagnostic on the sash position switches?


1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the Diagnostic

Diagnostic

option is

highlighted (it will turn white).

5.

Press

OK/Mute

to enter the Diagnostic submenu.





6.

Press the

button until the Diagnostic

Sash sensors

option is

highlighted (it will turn white). Press

OK/Mute

.





7.

Lower the sash, and note the status line – it should properly

report the sash as open too high, or open, or closed when the

sash is in the corresponding position.* Press

OK/Mute

to exit

the diagnostic.






*Early production models of Type A Logic+ units may display a sash height alarm

screen when the sash is raised above its operating position. This is normal.

MAIN MENU

1. MyLogic

TM

2. Settings

3. Service

 

or

to choose,

OK to select, MENU to escape

or

to choose,

OK to select, MENU to return

Current Settings

Restore Default

Calibration
Configuration

Diagnostics

 

Service

or

to choose,

OK to select, MENU to escape

Sash sensors

Keypad -Relays
Canopy Alarm
Airflow Sensor

 

Diagnostics

OK to return

Lower the sash from full open
to closed
Note the sash status below:
Sash Status:

The sash is open

The sash is open too high

The sash is closed

Diagnostics