Type B2 Exhaust HEPA Filter Leak Test

All Logic+ models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Logic+ have been appropriately

decontaminated before proceeding.

Note: If so equipped, the prefilter element on the towel catch should be

removed, and the aerosol generator should be placed nearby to inject the

aerosol as shown in Figure 5-15.

Figure 5-15








Testing at Labconco has shown the actual concentration varies from the calculated

value. Table 6-2 shows the actual vs. calculated concentrations for the B2 models.

Table 5-3

Type B2 Exhaust HEPA Filter Leak Test Specifications

Model

30340

30360

Air Displacement (CFM)

723

1083

Laskin Nozzles needed

2

2

Theoretical aerosol conc.

(ug/l)

1

38

25

Actual aerosol conc. (ug/l)

1

21

17

3.

The calculated concentration was established by the formula:

Concentration (ug/l) = (# Laskin nozzles @ 20 PSIG x 13,500)/Volume of air displaced

4.

Based on Mineral oil

1.

Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2.

Check the oil level of the generator – it must be within 1/8” of the level line

when the generator is sitting level.

3.

Establish the 100% and 0% concentration levels for the photometer.