Inflow velocity calculation - secondary method – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 88

Advertising
background image

Product Service 1-800-522-7658

88

Inflow Velocity Calculation - Secondary

Method

If the primary method cannot be performed, the secondary inflow calculation method

should be used. The secondary method uses a hot wire thermal anemometer. In this

method, the sash is lowered, and a series of velocity readings are taken at the front of

the unit.
NOTE: You must use Labconco holder P/N 3836405 to perform this test

properly. Qualified certifiers can obtain this holder as part of the Logic+

Certifier Kit, Labconco P/N 3858403, by contacting Labconco Customer

Service Department at 800-821-5525.

For each model Purifier, insert the proper template as indicated in Table 5-6, and shown

below in Figure 5-18.

The thermal anemometer probe should be positioned such that the distance from the

bottom of the probe holder to the center of the sensor element equals “Sensor Distance”

as shown in Table 5-6. Use the marks on the 6-foot templates, or a scale as shown in

Figure 5-19 & 5-20 below.

Figure 5-18


Measuring the restricted access opening

height, using the appropriate template.

Ensure that the bottom of the template is

seated in the sash stop.