Labconco PURICARE Procedure Station Models 31260 User Manual

Page 107

Advertising
background image

Product Service 1-800-522-7658

107

6.

Using a 1/2 -inch socket or wrench, loosen the four exhaust filter sealing bolts

located at each corner of the exhaust filter tray, as shown in Figure 6-17, at least

1/2 inch. This releases the exhaust HEPA filter from the cabinet.

Figure 6-17


Exhaust HEPA Filter bolts

(2 per each side)






7.

Pull the exhaust HEPA filter straight out of the cabinet.

To Remove the Exhaust HEPA Filter on 3-foot models:

1.

Follow steps 1-5 as above.

2.

Loosen the four sealed exhaust filter bolts on the top of the cabinet hull, as

shown in Figure 6-18 and 6-19.

3.

Remove the exhaust filter by pulling it straight out of the cabinet.

Figure 6-18



Exhaust HEPA Filter bolts