Labconco PURICARE Procedure Station Models 31260 User Manual

Page 144

Advertising
background image

Product Service 1-800-522-7658

144

9.

Note the orientation of the sensor

harness plug and the sensor; it is a

keyed connection-and can only be

connected in the correct orientation.

Plug the sensor harness into the

sensor, and the other end into the

multi pin connector on the Logic+’s

electronics module.

10.

Reconnect the Logic to electrical service. The unit is now ready for the calibration

of the air flow sensor.

Note: You will need to access the certifiers menu to calibrate the airflow

sensor.

Note: To properly calibrate the air flow sensor, a qualified certifier must

verify the inflow and downflow values.

With the Logic operating:

1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted

(it will turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.

4.

Press the

button until the

Calibration

option is

highlighted (it will turn white).

5.

Press

OK/Mute

to start the calibration procedure.

6.

You will see a

warning screen

alerting you that you are

about to alter the BSC’s settings.

7.

Press

OK/Mute

to input the password.


8.

When requested for the password press

Light

,

UV light

,

Timer

,

Timer

then

OK/Mute

.

or

to choose,

OK to select, MENU to return

Current Settings

Restore Default

Diagnostics

Configuration

Calibration

 

Service

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Calibration

OK to proceed, MENU to return

PASSWORD:

Calibration

MAIN MENU

1. MyLogic

TM

2. Settings

3. Service

 

or

to choose,

OK to select, MENU to escape