Type a hepa filter leak test specifications, Table 5-1 – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 73

Advertising
background image

Product Service 1-800-522-7658

73

Table 5-1

Type A HEPA Filter Leak Test Specifications

Cabinet

Width (Feet)

3

3

4

4

5

5

6

6

Sash Height

(inches)

10, 12

8, 9

10, 12

8, 9

10, 12

8, 9

10, 12

8, 9

Air

Displacement

(CFM)

596

543

794

723

991

903

1189

1083

Laskin Nozzles

needed

1

1

2

2

2

2

2

2

Calculated

aerosol conc.

(ug/l)

1

23

25

34

38

27

30

23

25

Actual aerosol

conc. (ug/l)

2

12

14

18

21

15

17

15

17

1.

The calculated concentration was established by the formula:

Concentration (ug/l) = (# Laskin nozzles @ 20 PSIG x 13,500)/Volume of air displaced

2.

Based on Mineral oil

1.

Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2.

Check the oil level of the generator – it must be within 1/8” of the level line

when the generator is sitting level.

3.

Establish the 100% and 0% concentration levels for the photometer.