Product Service 1-800-522-7658

106

Figure 6-14

Figure 6-15

5.

Remove the last two screws. Gently rock the cover, slide it to the left or right, and

swing it out and away from the unit.

NOTE: During disassembly, the blower plenum cover can rest on the center sash

support, as shown in Figure 6-16, to allow repositioning of the panel or your grip on it.

Figure 6-16

Center Sash Support


6.