Electronics module access, Electronic module component identification – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 124

Advertising
background image

Product Service 1-800-522-7658

124

Electronics Module Access

The Electronics module is located on the top right side of the cabinet. To access the

module, do the following:

1.

Unplug the Logic+, either at the wall, or at the power cord inlet at the back of

the electronics module, as shown in Figure 7-2.

2.

Remove the three 7/16-inch acorn nuts and lock washers that secure the

module.

3.

Tip the module on its side, and secure it so it does not fall.

Figure 7-2



Electronic Module Component Identification

The Electronics module contains the circuit breakers, the relay and power supply boards,

the fluorescent and optional UV ballasts, and the optional RS-232 board. These items

are shown in Figure 7-3.

Figure 7-3

Circuit Breakers

Fluorescent Ballast

UV Ballast

Outlet Relay

Light Relay

UV Light Relay

Power Supply Circuit Board

Advertising