The Logic+ has three circuit boards. The ECM microprocessor and power

supply board, located in the end of the ECM, the control board on the right

side wall, shown in Figure 2-5, and the power supply board assembly,

located in the electronics module, as shown in Figure 1-13.

Figure 1-13

UV Relay


Light Relay

Outlet Relay




The Logic+ also has an optional contact/RS232 board. This must be installed

if the user wants to signal external building monitors or controllers, or wants

the Logic+ to output operating information to a data logger. The optional

board is located in the electronics module, shown in Figure 1-14.