Type A Supply HEPA Filter Test

Downflow Diffuser Removal

1.

Remove the diffuser by raising the sash as high as it will go, and then removing

the two acorn nuts with a 7/16” wrench or socket, as shown in Figure 5-3.

Figure 5-3

Removing the diffuser acorn nuts (Dress

panel and lamps removed for clarity

only).


Note: The diffuser acorn nuts are secured to the studs with a removable thread locking

compound to prevent them from vibrating loose during shipment of the Logic+. There

may be some resistance the first time the nuts are removed; this is normal. No

additional thread locking compound need be applied to these acorn nuts unless the unit

is being prepared for shipment.

2.

Lower the front edge of the diffuser until clears the bottom of the sash, and then

pull the diffuser straight out of the cabinet, as shown in Figure 5-4.

Figure 5-4


3.

The Logic+ is now ready for the HEPA Filter Leak Test. Set the photometer

sampling valve to “DOWNSTREAM”. Ensure there is a proper vacuum at the

sampling port of the pistol.