Table 5-5

Inflow Test Specifications – 8 inch Sash Opening

Width (feet)

3

4

5

6

Nominal Average Inflow

(FPM)

105+/-5

105+/-5

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

2.03

2.69

3.36

4.03

Nominal Avg. Exhaust Vol.

(CFM)

213

283

353

423

Avg. Exhaust Vol. Range

(CFM)

203-223

269-296

336-370

403-443

Inflow Test Specifications – 9 inch Sash Opening

Width (feet)

3

4

5

6

Nominal Average Inflow

(FPM)

105+/-5

105+/-5

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

2.28

3.03

3.78

4.53

Nominal Avg. Exhaust Vol.

(CFM)

240

318

397

476

Avg. Exhaust Vol. Range

(CFM)

228-251

303-333

378-416

453-498

Inflow Test Specifications – 10 inch Sash Opening

Width (feet)

3

4

5

6

Nominal Average Inflow

(FPM)

105+/-5

105+/-5

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

2.53

3.37

4.2

5.03

Nominal Avg. Exhaust Vol.

(CFM)

266

354

441

529

Avg. Exhaust Vol. Range

(CFM)

253-278

337-371

420-462

503-553

Inflow Test Specifications – 12 inch Sash Opening

Width (feet)

3

4

5

6

Nominal Average Inflow

(FPM)

85+/-5

85+/-5

85+/-5

85+/-5

Sash Open Area (Sq. Ft)

3.04

4.04

5.04

6.04

Nominal Avg. Exhaust Vol.

(CFM)

259

344

428

514

Avg. Exhaust Vol. Range

(CFM)

243-274

323-364

403-454

483-544


Acceptance

Average inflow shall be meet the specified value.

Advertising