Uv settings, Uv lamp hourmeter, Reset uv lamp hourmeter – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 68: Change uv lamp life

Advertising
background image

Product Service 1-800-522-7658

68

UV Settings

For models equipped with the optional UV light the Logic+ has an integral UV light

maintenance system. It allows you to define how many hours you want the UV lamp to

operate before receiving a reminder to replace it, a way to monitor how many hours the

lamp has been on, and the means to reset the UV lamp hourmeter.

UV Lamp Hourmeter

This display only shows how many hours the UV lamp has been lit, and how many hours

remain until you will receive a warning to replace the lamp.

From the Settings menu screen, use the

and

buttons on the touchpad to highlight the

UV Settings

option-it will turn white when selected. Press

OK/MUTE

to

enter the UV Settings screen. Using the

and

buttons

on the touchpad, highlight UV Hourmeter-it will turn white

when selected, and then press

OK/Mute

. The screen will

then return to the UV Hourmeter screen.

This screen displays how many hours the UV lamp has

operated, and how many hours of operation remain before

replacement is recommended. When finished with this

screen, press

MENU

to return to the first UV Settings

screen.

Reset UV Lamp Hourmeter

This option lets you reset the UV hourmeter to 0 hours

whenever the lamp has been replaced.

This screen allows you to reset the UV Hourmeter to 0

hours if you press

OK/Mute

. Pressing

MENU

will return

you to the first UV Settings screen without resetting the

hourmeter.

Change UV Lamp Life

In this screen, you can set the number of operating hours before receiving the replace

UV lamp warning. For most UV lamps, the output of UV light decreases at a constant

rate. Typically, after 6,000 hours of operation the lamp will output 80%. This option

allows you to set operational life of the UV lamp, in 100 hour increments.

From the UV Settings menu screen, use the

and

buttons on the touchpad to highlight the

Set UV Life

option-

it will turn white when selected. Press

OK/MUTE

to enter

the Set UV life screen. Use the

and

buttons to increase

or decrease the lamp life in 100 hour increments, until the

desired value is displayed, and then press

OK/Mute

. The

screen will then return to the UV Settings screen.