Lamp Socket Replacement

1.

Unplug the Logic+

2.

Remove the dress panel as noted in Figure 6-1

3.

Remove the fluorescent lamps from the sockets.

4.

To replace a socket, insert a small straight bladed screwdriver into the socket

near the lamp wire, as shown in Figure 7-1.

5.

Press the wire into the new socket.

6.

Reassemble the unit.

Figure 7-1

Optional Ultraviolet Lamp

Lamp Identification

All Logic+s use a single 30-Watt germicidal lamp. The lamp information is shown

below. The code of the lamp part number is as follows:

G30T8

G = germicidal lamp

30 = the lamp Wattage

T8 = the lamp mounting configuration

UV lamp Specifications

For all models, the UV lamp number is G30T8. The Labconco part number

is 1271300.

For all 115/230 VAC models, the ballast assembly is Labconco part

number 3829901. The ballast is Robertson part number

PSM2GPH18MVW.

For all 100 VAC models, the ballast assembly is Labconco part number

3830600. The ballast is Robertson part number RSO1GPH30100.