Preparing the Logic+ for Gaseous

Sterilization

Note: This section only reviews the steps required for preparing the Logic+

for Gas Sterilization. Thoroughly understand the sterilization procedures and

protocols supplied by the manufacturer of the sterilizing system before

attempting this operation.

1.

Unplug the Logic+ from electrical power.

2.

Thoroughly surface decontaminate the working area of the cabinet.

In order to get a complete seal on the front of the cabinet, do the following:

1.

Remove the work surface after it has been surface decontaminated.

2.

Remove the towel catch, and locate the Sampling/Decontamination Tube, as

shown in Figure 6-32.

3.

Remove the front grille.

4.

Place the work surface and the grille into the work area for decontamination.

5.

Remove the front dress panel.

6.

Remove the sash stop bracket, located near the top of the blower plenum cover.

7.

Lift the sash up until the bottom edge of the sash clears the bottom edge of the

Supply HEPA Filter shelf.

8.

Using flexible plastic sheet 5-8 mil thick and duct tape seal the perimeter of the

front opening. If you wish to connect the sterilizer’s discharge tube to the
Logic+’s Sampling/Decontamination Tube, do this now, and properly seal the

discharge tube to the plastic sheet.

9.

Open the exhaust filter cover.

Note: DO NOT attempt to remove the brackets that secure the exhaust filter cover;

they are needed to maintain rigidity on the top of the chassis.
10.

Using flexible plastic sheet and duct tape seal the Exhaust filter opening.

11.

The Logic+ is now prepared for gaseous decontamination.