Labconco PURICARE Procedure Station Models 31260 User Manual

Page 165

Advertising
background image

Product Service 1-800-522-7658

165


Relay Contact #2……the “airflow error” contact

This set of contacts (dry contacts pins 1 & 2 of J3 on the electrical enclosure, and the

ORANGE wires on Labconco harness P/N 3432300) will indicate when the Logic+ has

detected an airflow error, as described by the conditions below:

Type of Logic+
unit

Alarm
Condition

LCD display
message

Airflow Error
dry contacts

Comments

All Labconco
Logic+ Cabinets

Sash out of
position

SASH TOO HIGH

No Response

Motor RPM has

shifted abruptly

AIRFLOW ALERT

Contacts shorted

HEPA Filter
approaching end
of life—20%

CHECK FILTER
(Filter life indicator
at 20% or less)

No response

HEPA Filter

approaching end
of life-10%

CHECK FILTER

(Filter life indicator
at 10% or less)

Contacts shorted

HEPA Filter end

of life- 0%

CHECK FILTER

(Filter life indicator
at 0%)

Contacts shorted

Unit beeps

approximately
every 15 seconds

**Note all
specialty Logic+
BSC’s below will

have all the
alarm conditions

listed above, plus
specific alarms
listed below”

A2 BSC with
canopy sensor**

( see above)

Canopy Airflow
incorrect

CANOPY ERROR

Contacts shorted

A2 BSC with

airflow sensor**(
see above)

Downflow Airflow

incorrect

Alarm - Airflow

Contacts shorted

B2 BSC**( see
above)

Exhaust airflow
not correct

Alarm - Inflow

Contacts shorted