Figure 5-17



















Acceptance

Average downflow shall be 55 +/-5 FPM with all test points being within +/- 16 FPM.


Inflow Velocity Calculation - Primary
Method

Note: The dress panel MUST be installed on the cabinet to obtain an accurate inflow

value.
The average inflow velocity for Logic+ A2 and B2 models is 105 ±5 FPM, and Type A1s

are 85+/-5 FPM. It should be determined by converting the inflow volume to the

average inflow velocity. The inflow volume is measured directly by sealing a direct inflow

measuring flow meter to the face of the unit. When corrected for local conditions,

dividing the inflow volume by the opening area will yield the average inflow velocity.