Labconco PURICARE Procedure Station Models 31260 User Manual

Page 169

Advertising
background image

Product Service 1-800-522-7658

169

exhaust outlet

not blocked or restricted.

Blockage or

restriction under

the work surface

Ensure that the towel catch and

plenum beneath the work surface

are unobstructed.

Contamination of

work in the

cabinet

Improper technique

or procedure for

the Biosafety

Cabinet

See “Use of the Cabinet” section in

the manual.

Restriction of the

return air slots or

grille – blockage of

the exhaust outlet

Ensure that all return air slots,

grilles and the exhaust outlet are

unobstructed.

External factors are

disrupting the

cabinet airflow

patterns or acting

as a source of

contamination

See “Installation” section of this

manual.

Cabinet is out of

adjustment/HEPA

filter(s) are

defective

Have cabinet recertified.