4.9. SETUP PASS: PASSWORD FEATURE

The 9110TH/M provides password protection of the calibration and setup functions to
prevent unauthorized adjustments. When the passwords have been enabled in the PASS
menu item, the system will prompt the user for a password anytime a password-
protected function (e.g., SETUP) is selected. This allows normal operation of the
instrument, but requires the password (101) to access to the menus under SETUP. When
PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup
(SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is
enabled or disabled, a password (default 818) is required to enter the VARS or DIAG
menus in the SETUP>MORE menu.

There are three levels of password protection, which correspond to operator,
maintenance, and configuration functions. Each level allows access to all of the
functions in the previous level.

Table 4-10: Password Levels

Password Level

Menu

Access

Allowed

Null (000)

Operation

All functions of the MAIN menu: TEST, GEN, initiate SEQ , MSG, CLR

101 Configuration/Maintenance Access to Primary Setup and Secondary SETUP Menus when

PASSWORD is enabled.

818 Configuration/Maintenance Access to Secondary SETUP Submenus VARS and DIAG whether

PASSWORD is enabled or disabled.

To enable or disable passwords, press the following menu button sequence:

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE

PASS

CLK MORE

EXIT

Toggle this

button to

enable, disable

password

feature

SETUP X.X

PASSWORD ENABLE: OFF

OFF

ENTR EXIT

SETUP X.X

PASSWORD ENABLE: ON

ON

ENTR

EXIT