Model 9110TH NOx Analyzer

Troubleshooting & Repair

Teledyne Analytical Instruments

252

7.7. REMOVING / REPLACING THE RELAY PCA FROM THE

INSTRUMENT

This is the most commonly used version of the Relay PCA. It includes a bank of solid
state AC relays. This version is installed in analyzers where components such as AC
powered heaters must be turned ON & OFF. A retainer plate is installed over the relay
to keep them securely seated in their sockets.

AC Relay

Retainer Plate

Retainer

Mounting

Screws

Figure 7-20:

Relay PCA with AC Relay Retainer In Place

The Relay retainer plate installed on the relay PCA covers the lower right mounting
screw of the relay PCA. Therefore, when removing the relay PCA, the retainer plate
must be removed first.

AC Relay Retain Occludes

Mounting Screw on

P/N 045230200

Mounting

Screws

Figure 7-21:

Relay PCA Mounting Screw Locations