To define the REPORT PERIOD, follow the instruction shown in Section 4.7.2.2 then
press:


Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 8, 8500

PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu.

SETUP X.X

NAME: CONC

<SET

SET>

EDIT PRINT

EXIT

Press SET> until you reach REPORT PERIOD (OR SAMPLE PERIOD) …

SETUP X.

X REPORT PERIODD:DAYS:0


0 0 0

ENTR EXIT

ENTR accepts the new string and

returns to the previous menu.

EXIT ignores the new string and

returns to the previous menu.

Press buttons to set hours

between reports in the format :
HH:MM (max: 23:59). This is a

24 hour clock . PM hours are 13

thru 23, midnight is 00:00.
Example 2:15 PM = 14:15

Set the number of days

between reports (0-366).

IIf at any time an illegal entry is selected (e.g., days > 366)

the ENTR button will disappear from the display.

SETUP X.X

REPORT PERIODD:TIME:01:01

0 1 0 0

ENTR EXIT

Use the PREV and NEXT

buttons to scroll to the

data channel to be edited.

SETUP X.X

REPORT PERIOD:000:01:00


<SET SET>

EDIT

PRINT

EXIT

From the DATA ACQUISITION menu

(see Section 6.7.2.2)

SETUP X.X

ENTER DAS PASS: 818

9 2 9

ENTR

EXIT

Changing the SAMPLE

PERIOD or REPORT

PERIOD Requires a

special password