Relay board and status leds – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 228

Advertising
background image

Model 9110TH NOx Analyzer

Troubleshooting & Repair

Teledyne Analytical Instruments

208

7.1.4.3. Relay Board and Status LEDs

The most important status LED on the relay board is the red I

2

C Bus watch-dog LED,

labeled D1, which indicates the health of the I

2

C communications bus. This LED is the

left-most in LED row 1 in the center of the relay board when looking at the electronic
components. If D1 is blinking, then the other LEDs can be used in conjunction with the
DIAG menu I/O functions to test hardware functionality by manually switching devices
on and off and watching the corresponding LED go on or off.

Figure 7-4 illustrates the relay board layout including the two rows of LEDs,

Table 11-2 lists the individual LED functions and the menu tree below shows how to
access the manual control of the I/O functions that only some or the LEDs may be
functional in your analyzer model; the relay board layout is conceptualized for spare,
future functionality and is also common to many of the E-series analyzers.

Power

Connection

for DC

Heaters

Status LED’s

(D2 through D16)

DC Power Supply

Test Points

Watchdog

Status LED (D1)

(JP5)

Thermocouple

Configuration

Jumpers

Thermocouple

Signal Output

I

2

C

Connector

Shutter Control

Connector

(T100 Series Only)

Valve Control

Drivers

Pump Power

Output

(JP7)

Pump AC

Configuration

Jumper

AC Power

IN

AC Heater

Power Output

AC Power Output for
Optional O

2

sensors

(JP6)

Main AC Heater

Configuration Jumpers

(JP2)

AC Configuration Jumpers

for Optional IZS Valve

Heaters & O

2

Sensors

Solid State AC

Power Relays

(Not Present on
P/N 45230100)

DC Power

Distribution

Connectors

Valve Option

Control

Connector

(J15)

TC1 Input

(J16)

TC2 Input

Figure 7-4:

Relay Board PCA