Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

30

VACUUM

PRESSURE

SENSOR

SAMPLE

PRESSURE

SENSOR

O3

FLOW

SE

N

S

OR

FLOW PRESSURE

SENSOR PCA

INSTRUMENT CHASSIS

PERMAPURE
DRYER

NO/NO

X

VALVE

AUTOZERO

VALVE

O

3

GENERATOR

SAMPLE

GAS

INLET

Fil
ter

Orifice Dia.

0.004"

PUMP

NO

X

E

x

ha

us

t

Scr

u

b

b

er

O

3

Destruct

EXHAUST

GAS

OUTLET

EX

HA

US

T M

A

NIF

O

LD

Orifice Dia.

0.007"

Orifice Dia.

0.007"

REACTION

CELL

PMT

NC

COM

NO

NC

COM

NO

NO

2

Converter

Figure 3-16:

9110TM Internal Pneumatic Block Diagram - Standard Configuration

Note

Pneumatic Diagrams do not reflect the physical layout of the instrument.

The most significant differences between the 9110TH and 9110TM versions in regards
to pneumatic flow are:

 A bypass line leading directly from the particulate filter to the exhaust

manifold is present on the 9110TH, but not in the 9110TM.

 The diameter of the critical flow orifice controlling the gas flow into the

sample chamber is smaller and therefore the flow rate of sample gas through
the instrument is lower.