Pmt cooling system, Pmt preamplifier, Tec control board – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 304: And no

Advertising
background image

Model 9110TH NOx Analyzer

Principles

of

Operation

Teledyne Analytical Instruments

284

8.5.4. PMT COOLING SYSTEM

The performance of the analyzer’s PMT is significantly affected by temperature.
Variations in PMT temperature are directly reflected in the signal output of the PMT.
Also the signal to noise ratio of the PMT output is radically influenced by temperature
as well. The warmer The PMT is, the noisier its signal becomes until the noise renders
the concentration signal useless. To alleviate this problem a special cooling system
exists that maintains the PMT temperature at a stable, low level

TEC

Control

PCA

PMT Preamp

PCA

Thermistor

outputs temp of

cold block to

preamp PCA

Preamp PCA sends

buffered and

amplified thermistor

signal to TEC PCA

TEC PCA sets

appropriate

drive voltage

for cooler

Heat form PMT is absorbed

by the cold block and

transferred to the heat sink

via the TEC then bled off

into the cool air stream.

PMT

Cold Block

Hea

t Si

n

k

Cooling Fan

ThermoElectric Cooler

Figure 8-19:

PMT Cooling System

8.5.4.1. TEC Control Board

The TEC control printed circuit assembly is located in the sensor housing assembly,
under the slanted shroud, next to the cooling fins and directly above the cooling fan.
Using the amplified PMT temperature signal from the PMT preamplifier board (see
Section 10.4.5), it sets the drive voltage for the thermoelectric cooler. The warmer the
PMT gets, the more current is passed through the TEC causing it to pump more heat to
the heat sink.

A red LED located on the top edge of this circuit board indicates that the control circuit
is receiving power. Four test points are also located at the top of this assembly. For the
definitions and acceptable signal levels of these test points see Section 11.

8.5.5. PMT PREAMPLIFIER

The PMT preamplifier board amplifies the PMT signal into a useable analog voltage
(PMT) that can be processed by the motherboard into a digital signal to be used by the
CPU to calculate the NO, NO

2

and NO

x

concentrations of the gas in the sample

chamber.

The output signal of the PMT is controlled by two different adjustments. First, the
voltage across the electron multiplier array of the PMT is adjusted with a set of two
hexadecimal switches. Adjusting this voltage directly affects the HVPS voltage and,
hence, the signal from the PMT. Secondly, the gain of the amplified signal can further