7. TROUBLESHOOTING & REPAIR

This section contains a variety of methods for identifying and solving performance
problems with the analyzer.

CAUTION

The operations outlined in this Section must be performed by qualified

maintenance personnel only.

WARNING

Risk of electrical shock. Some operations need to be carried out with

the analyzer open and running. Exercise caution to avoid electrical
shocks and electrostatic or mechanical damage to the analyzer. Do not

drop tools into the analyzer or leave those after your procedures. Do
not shorten or touch electric connections with metallic tools while
operating inside the analyzer. Use common sense when operating

inside a running analyzer.

7.1. GENERAL TROUBLESHOOTING

The analyzer has been designed so that problems can be rapidly detected, evaluated and
repaired. During operation, the analyzer continuously performs diagnostic tests and
provides the ability to evaluate its key operating parameters without disturbing
monitoring operations.

A systematic approach to troubleshooting will generally consist of the following five
steps:

 any warning messages and take corrective action as necessary.

 Examine the values of all TEST functions and compare them to factory

values. any major deviations from the factory values and take corrective
action.

 Use the internal electronic status LED’s to determine whether the electronic

communication channels are operating properly. Verify that the DC power
supplies are operating properly by checking the voltage test points on the
relay board. that the analyzer’s DC power wiring is color-coded and these
colors match the color of the corresponding test points on the relay board.