Perform zero/span calibration, Step three – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 200

Advertising
background image

Model 9110TH NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

180

STEP THREE:

Perform Zero/Span Calibration:

Press ENTR to changes

the OFFSET & SLOPE

values for both the NO

and NO

x

measurements.

Press EXIT to leave the

calibration unchanged and

return to the previous

menu.

SAMPLE

GAS TO CAL:NOX

NOX

O2

ENTR

EXIT

ZERO CAL M

NOX STB= XXX.X PPM

NOX=XXX.X

<TST TST>

ZERO

CONC

EXIT

SAMPLE

RANGE TO CAL:LOW

LOW

HIGH

ENTR

EXIT

EXIT at this point

returns to the

SAMPLE menu.

Press ENTR to changes

the OFFSET & SLOPE

values for both the NO

and NO

x

measurements.

Press EXIT to leave the

calibration unchanged and

return to the previous

menu.

Set the Display to show

the NOX STB test

function.

This function calculates

the stability of the NO/NO

x

measurement

Toggle TST> button until ...

Allow zero gas to enter the sample port

at the rear of the analyzer.

Wait until NOX STB
falls below 0.5 ppm.

This may take several

minutes.

ZERO CAL M

NOX STB= XXX.X PPM

NOX=XXX.X

<TST TST>

ENTR

CONC

EXIT

SAMPLE

GAS TO CAL:NOX

NOX

O2

ENTR

EXIT

SPAN CAL M

NOX STB= XXX.X PPM

NOX=X.XXX

<TST TST> ZERO

SPAN

CONC

EXIT

SAMPLE

RANGE TO CAL:LOW

LOW

HIGH

ENTR

EXIT

Allow span gas to enter the sample port

at the rear of the analyzer.

Wait until NOX STB
falls below 0.5 ppm.

This may take several

minutes.

SPAN CAL M

NOX STB= XXX.X PPM

NOX=X.XXX

<TST TST>

ENTR

CONC

EXIT

SPAN CAL M

NOX STB= XXX.X PPM

NOX=X.XXX

<TST TST> ENTR

CONC

EXIT

Analyzers enters SPAN cal

mode and the SPAN key

appears.

You may see both

keysduring the transition

from ZERO to SPAN modes.

If either the ZERO or SPAN

buttons fail to appear see

Section 11 for

troubleshooting tips.

Analyzer continues to

cycle through NO

x

,

NO, and NO

2

measurements
throughout this

procedure.

SAMPLE

A1:NXCNC1=100PPM

NOX=XXX.X

< TST

TST >

CAL CALZ

CALS

SETUP

SAMPLE

NOX STB= XXX.X PPM

NOX=XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

SAMPLE

NOX STB= XXX.X PPM

NOX=XXX.X

< TST TST > CAL CALZ

CALS

SETUP

Analyzers enters

ZERO cal

mode.

SAMPLE

NOX STB= XXX.X PPM

NOX=XXX.X

< TST TST > CAL

CALZ CALS

SETUP