Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 138

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

118

5. Set the reporting range scale for the data type chosen.

6. Set the update rate for the channel.

7. Calibrating the output channel. This can be done automatically or manually for

each channel (see Sections 4.13.6).

To access the analog I/O configuration sub menu, press:

AIO Configuration Submenu

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

SETUP X.X

ENTER PASSWORD:818

8

1

8

ENTR

EXIT

DIAG SIGNAL I/O

NEXT

ENTR

EXIT

Continue pressing NEXT until ...

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

DIAG AIO A OUTS CALIBRATED: NO

<SET

SET>

CAL

EXIT

DIAG AIO DATA_OUT_1: 5V, NXCNC1, NOCAL

<SET

SET>

EDIT

EXIT

DIAG AIO AIN CALIBRATED: NO

<SET

SET>

CAL

EXIT

DIAG AIO DATA_OUT_2: 5V, NXCNC1, NOCAL

<SET

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL

<SET

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_4: 5V, NXCNC1, NOCAL

<SET

SET>

EDIT

EXIT