Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 70

Advertising
background image

Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

50

STEP 2 – Activate O

2

sensor stability function

To change the stability test function from NO

x

concentration to the O

2

sensor output,

press:

SETUP X.X

STABIL_GAS:O2

NO NO2 NOX O2

ENTR EXIT

Press ENTR to keep

changes, then press

EXIT 3 times to return

to SAMPLE menu

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG ALRM

EXIT

SETUP X.X

ENTER PASSWORD:818

8

1

8

ENTR

EXIT

SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes

<PREV

NEXT>

JUMP

EDIT PRNT EXIT

SETUP X.X

2) STABIL_GAS=NOX

<PREV NEXT> JUMP

EDIT

PRNT EXIT

SETUP X.X

STABIL_GAS:NOX

NO NO2 NOX

O2

ENTR EXIT

Continue pressing NEXT until ...

Note

Use the same procedure to reset the STB test function to NO

x

when the O

2

calibration procedure is complete.