Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 70
Model 9110TH NOx Analyzer
Getting Started
Teledyne Analytical Instruments
50
STEP 2 – Activate O
2
sensor stability function
To change the stability test function from NO
x
concentration to the O
2
sensor output,
press:
SETUP X.X
STABIL_GAS:O2
NO NO2 NOX O2
ENTR EXIT
Press ENTR to keep
changes, then press
EXIT 3 times to return
to SAMPLE menu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG ALRM
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
2) STABIL_GAS=NOX
<PREV NEXT> JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS:NOX
NO NO2 NOX
O2
ENTR EXIT
Continue pressing NEXT until ...
Note
Use the same procedure to reset the STB test function to NO
x
when the O
2
calibration procedure is complete.