Analog output step test – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 133
Operating Instructions
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
113
4.13.3. ANALOG OUTPUT STEP TEST
This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the
operation of the data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG AOUT
ANALOG OUTPUT
0%
EXIT
Performs
analog output
step test.
0% - 100%
DIAG AOUT
ANALOG OUTPUT
[0%]
EXIT
Pressing the key under “0%” while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.