Analog output step test – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 133

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

113

4.13.3. ANALOG OUTPUT STEP TEST

This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the
operation of the data logging/recording devices attached to the analyzer.

To begin the Analog Output Step Test press:


SAMPLE A1:NXCNC1=100PPM NOX=XXX.X


< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

DIAG

SIGNAL I / O


NEXT

ENTR

EXIT

Exit-Exit

returns to the

DIAG menu

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG AOUT

ANALOG OUTPUT

0%

EXIT

Performs

analog output

step test.

0% - 100%

DIAG AOUT

ANALOG OUTPUT

[0%]

EXIT

Pressing the key under “0%” while performing the test will

pause the test at that level. Brackets will appear around

the value: example: [20%] Pressing the same key again

will resume the test.