Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

136

 As before, adjustment of the output is performed until the current reading of

the meter reaches the desired point (e.g. 2 mA, 4 mA, 20 mA, etc.)

mA

IN OUT

I OUT +

I OUT -

I IN +

I IN -

Recording
Device

Analyzer

See Table 3-2 for

pin assignments of

the Analog Out

connector on the

rear panel.

Current

Meter

Figure 4-12:

Setup for Calibrating Current Outputs

Note

Do not exceed 60 V between current loop outputs and instrument ground.

If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250

 1% resistor across the current loop output. Using a

voltmeter, connected across the resistor, follow the procedure above but adjust the
output to the following values:

V

+DC Gnd

Recording

Device

V IN +

V IN -

ANALYZER

V OUT +

V OUT -

250 O

Volt

Meter

Figure 4-13:

Alternative Setup for Calibrating Current Outputs