Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 156
Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
136
As before, adjustment of the output is performed until the current reading of
the meter reaches the desired point (e.g. 2 mA, 4 mA, 20 mA, etc.)
mA
IN OUT
I OUT +
I OUT -
I IN +
I IN -
Recording
Device
Analyzer
See Table 3-2 for
pin assignments of
the Analog Out
connector on the
rear panel.
Current
Meter
Figure 4-12:
Setup for Calibrating Current Outputs
Note
Do not exceed 60 V between current loop outputs and instrument ground.
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250
1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the
output to the following values:
V
+DC Gnd
Recording
Device
V IN +
V IN -
ANALYZER
V OUT +
V OUT -
250 O
Volt
Meter
Figure 4-13:
Alternative Setup for Calibrating Current Outputs