The default settings for alm1 and alm2 are – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 168

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

148

4.14. SETUP – ALRM: USING THE OPTIONAL GAS

CONCENTRATION ALARMS (OPT 67)

The optional alarm relay outputs (Option 67) are installed includes two concentration
alarms. Each alarm has a user settable limit, and is associated with an opto-isolated TTL
relay accessible via the status output connector on the instrument’s back panel. If the
concentration measured by the instrument rises above that limit, the alarm‘s status
output relay is closed NO

2

.

The default settings for ALM1 and ALM2 are:

Table 4-28: Concentration Alarm Default Settings

ALARM STATUS

LIMIT SET POINT

1

OUTPUT RELAY

DESIGNATION

ALM1

Disabled

100 ppm

133.9 mg/m

3

AL2

ALM2

Disabled

300 ppm

401.6 mg/m

3

AL3

1

Set points listed are for PPM. Should the reporting range units of measure be changed the analyzer will automatically

scale the set points to match the new range unit setting.

Note

To prevent the concentration alarms from activating during span
calibration operations make sure to press CAL or CALS button prior to
introducing span gas into the analyzer.

To enable either of the concentration alarms and set the Limit points, press:

ENTR accepts the new

settings

EXIT ignores the new

settings

SETUP X.

ALARM 1 LIMIT: 200 PPM

0 1 0 0 .0 0

ENTR EXIT

SETUP X.X

ALARM MENU

ALM1 ALM2

EXIT

SETUP X.

ALARM 1 LIMIT: OFF

OFF

ENTR EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

ALRM

EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

Toggle these buttons

to cycle through the

available character set:

0-9

SETUP X.

ALARM 1 LIMIT: ON

ON

ENTR EXIT