Automatic analog output calibration – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 152
Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
132
4.13.6.1. Automatic Analog Output Calibration
To calibrate the outputs as a group with the AOUTS CALIBRATION command, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL is turned
ON
If any of the channels
have not been calibrated
ot if at least one channel
has AUTO-CAL turned
OFF, this message will
read NO.
DIAG AIO
AUTO CALIBRATING DATA_OUT_1
DIAG AIO
AUTO CALIBRATING DATA_OUT_2
DIAG AIO
NOT AUTO CAL. DATA_OUT_3
DIAG AIO
AUTO CALIBRATING DATA_OUT_4
This message
appears when
AUTO-CAL is
Turned OFF for
a channel
DIAG AIO AOUTS CALIBRATED: YES
SET> CAL
EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Note
Manual calibration should be used for the 0.1V range or in cases where
the outputs must be closely matched to the characteristics of the
recording device.