4.13.6.1. Automatic Analog Output Calibration

To calibrate the outputs as a group with the AOUTS CALIBRATION command, press:

From the

AIO CONFIGURATION SUBMENU

(See Section 6.13.4.1)

Analyzer

automatically

calibrates all

channels for which

AUTO-CAL is turned

ON

If any of the channels

have not been calibrated

ot if at least one channel

has AUTO-CAL turned
OFF, this message will

read NO.

DIAG AIO

AUTO CALIBRATING DATA_OUT_1

DIAG AIO

AUTO CALIBRATING DATA_OUT_2

DIAG AIO

NOT AUTO CAL. DATA_OUT_3

DIAG AIO

AUTO CALIBRATING DATA_OUT_4

This message
appears when
AUTO-CAL
is

Turned OFF for

a channel

DIAG AIO AOUTS CALIBRATED: YES

SET> CAL

EXIT

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

Note

Manual calibration should be used for the 0.1V range or in cases where
the outputs must be closely matched to the characteristics of the
recording device.