Accessing the diagnostic features, Signal i/o – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 131

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

111

4.13.1. ACCESSING THE DIAGNOSTIC FEATURES

To access the DIAG functions press the following keys:


SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

From this point

forward, EXIT returns

to the

SECONDARY

SETUP MENU

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG

DISPLAY SEQUENCE CONFIG.

PREV

NEXT

ENTR

EXIT

DIAG

ELECTRICAL TEST

PREV

NEXT

ENTR

EXIT

DIAG

OZONE GEN OVERRIDE

PREV

NEXT

ENTR

EXIT

EXIT returns

to the main

SAMPLE

display

DIAG

OPTIC TEST

PREV

NEXT

ENTR

EXIT

DIAG

FLOW CALIBRATION

PREV

NEXT

ENTR

EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

At this point EXIT

returns

to the PRIMARY

SETUP MENU

4.13.2. SIGNAL I/O

The signal I/O diagnostic mode allows to review and change the digital and analog
input/output functions of the analyzer. See Appendix A-4 for a complete list of the
parameters available for review under this menu.

Note

Changes to signal I/O settings will remain in effect only until the signal I/O
menu is exited. Exceptions are the ozone generator override and the flow
sensor calibration, which remain as entered when exiting.