Setting the hessen protocol response mode, And com2, the type selection affects both ports, Text – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 181

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

161

and COM2, the TYPE selection affects both ports.

4.15.4.4. Setting The Hessen Protocol Response Mode

The Teledyne API’ implementation of Hessen Protocol allows the user to choose one of
several different modes of response for the analyzer.

Table 6-28: 9110TH/M Hessen Protocol Response Modes

MODE ID

MODE DESCRIPTION

CMD

This is the Default Setting. Reponses from the instrument are encoded as the traditional command format.
Style and format of responses depend on exact coding of the initiating command.

BCC

Responses from the instrument are always delimited with <STX> (at the beginning of the response, <ETX>
(at the end of the response followed by a 2 digit Block Check Code (checksum), regardless of the command
encoding.

TEXT

Responses from the instrument are always delimited with <CR> at the beginning and the end of the string,
regardless of the command encoding.

To Select a Hessen response mode, press:

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG ALRM

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE RANGE = 500.000 PPB SO2 =XXX.X

< TST TST > CAL

SETUP

SETUP X.X

HESSEN RESPONSE MODE :CMD

<SET

SET>

EDIT

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID

HESN

COM1

COM2

EXIT

SETUP X.X

HESSEN VARIATION: TYPE 1

SET>

EDIT

EXIT

SETUP X.X

HESSEN RESPONSE MODE :CMD

BCC TEXT EDIT

ENTR EXIT

Press to

change

response

mode.

ENTR accepts the new

settings

EXIT ignores the new

settings