4.11.4. ANALYZER ID

Each type of TAI analyzer is configured with a default ID code. The default ID code for
all 9110TH/M analyzers is either 0 or 200. The ID number is only important if more
than one analyzer is connected to the same communications channel such as when
several analyzers are on the same Ethernet LAN (see Section 4.11.7); in a RS-232
multidrop chain (see Section 4.11.9) or operating over a RS-485 network (see Section
4.11.6). If two analyzers of the same model type are used on one channel, the ID codes
of one or both of the instruments needs to be changed so that they are unique to the
instruments. To edit the instrument’s ID code, press:

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID

INET

COM1

COM2

EXIT

SETUP X.

MACHINE ID: 200 ID

0 2 0 0

ENTR EXIT

Toggle these buttons

to cycle through the

available character set:

0-9

ENTR button accepts the

new settings

EXIT key ignores the new

settings

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

The ID can be any 4 digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)