Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

137

Table 4-25: Current Loop Output Calibration with Resistor

FULL SCALE

VOLTAGE FOR 2-20 MA

(measured across 250Ω resistor)

VOLTAGE FOR 4-20 MA

(measured across 250Ω resistor)

0%

0.5 V

1.0 V

100%

5.0 V

5.0 V

To adjust the zero and span values of the current outputs, press:

EXAMPLE

EXAMPLE

DIAG AIO

DATA_OUT_2 CURR-Z:

13 mV

U100 UP10 UP DOWN DN10 D100

ENTR

EXIT

DIAG AIO DATA_OUT_2 CURR-S: 4866 mV

U100 UP10 UP DOWN DN10 D100

ENTR EXIT

From the

AIO CONFIGURATION SUBMENU

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

Continue pressing SET> until you reach the

output to be configured

DIAG AIO DATA_OUT_2: CURR, NXCNC1, NOCAL

<SET SET>

EDIT

EXIT

Continue pressing SET> until ...

DIAG AIO DATA_OUT_2 CALIBRATED:NO

<SET SET>

CAL

EXIT

DIAG AIO DATA_OUT_2 RANGE: CURR

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_2 CURR-Z: 0 mV

U100

UP10

UP DOWN DN10 D100 ENTR EXIT

DIAG AIO DATA_OUT_2 CURR-S: 5000 mV

U100 UP10 UP DOWN DN10

D100

ENTR EXIT

DIAG AIO DATA_OUT_2 CALIBRATED: YES

<SET SET> CAL

EXIT

Increase or decrease

the current output by

100, 10 or 1 counts.

The resulting change in

output voltage is

displayed in the upper

line.

Continue adjustments

until the correct current

is measured with the

current meter.

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT